The TMS mobile radiographic units family covers, in the variety of its models, all the specific needs of the radiographic examination at the patient’s bed, when he can’t reach the department of Radiology, from his bed in the Intensive Care Dept., from the Plaster Room, from the ER, from the Paediatrics Dept. or from the [...]

Availability: In stock

Product Description (longer)
Download documentation for this device

The TMS mobile radiographic units family covers, in the variety of its models, all the specific needs of the radiographic examination at the patient’s bed, when he can’t reach the department of Radiology, from his bed in the Intensive Care Dept., from the Plaster Room, from the ER, from the Paediatrics Dept. or from the Operating Theatre, in the post-operative phase.

Electrical data

Voltage    -    230Vac ±10% monophase standard

115Vac ±10% monophase

Frequency-   50/60Hz standard

Absorbed current    -   Values of current absorbed by the unit in the different operative conditions and in the two power supply values:

Operative condition  -  115 Vac / 50Hz    230 Vac / 50Hz

Stand By    -                     0,5A       0,5A

Radiographie   -             32A                20A

Line compensation-      Automatic

Line resistance-            <0,8Ω @230Vac

<0,4Ω @115Vac

Standard mains plug-  16A @230Vac

Isolation class     -         Class I with applied part type B

Functioning

User’s interface-Keyboard with LCD alphanumeric display, 2 lines X 20 characters for all the operative parameters and messages of possible faulty status. Service program for faults finding. Microprocessor management.

Languages- Italian, English, French, German, Spanish Selection by means of configuration programme (accessible by Service personnel only).

Radiography control- By hand-switch with extendible cable. Upon the ignition, the unit is in manual mode with default values.

Radiological data

Working technique   -    At 2 points with setting of kV and mAs

Focus in all the working conditions  -  1,5mm

Generator power in DC current (IEC 60601-1)-  4kW @100kV @230Vac

1,5kW @100kV @115Vac

Inverter frequency          –              22kHz

Max. frequency in high voltage-  44kHz

Max. ripple – <2% @100kV

Rise time  -

kV variation range –  40 ÷ 110kV in step of 1kV